Seria CETATEST 815 - seria testerów szczelności z czujnikiem różnicy ciśnienia
Pressure ranges:
-1 bar, +/- 5 mbar, +/- 50 mbar, 200 mbar, 1 bar, 6 bar, 10 bar, 20 bar, 30 bar, 400 bar (high pressure version), negative and positive gauge pressure ranges can be combined.
Measuring ranges:
± 500 Pa / ± 5,000 Pa (display resolution: 1 Pa, internal resolution: effectively 0.01 Pa)
Result units:
Pa, hPa, PSI, Pa/s, hPa/s, mbar • l/s, ml/min, ml/h, l/min, l/h, mmHg, mmWs, Torr
Types of test:
Standard: Pressure decay
Optional: sealed component, dynamic pressure measurement, pressure steps, pressure rise (with double leak connector)
Parameters can be exported to a USB stick or imported again. The storage of measurement series and waveforms is possible.
Interfaces:
Standard: Digital I / O, RS-232
Optional: Profibus DP, Profinet, Ethernet, EtherCAT, Bluetooth, USB
Dimensions:
345 mm x 145 mm x 435 mm (W x H x D)
In the high-pressure version, leak tests up to a test pressure of 400 bar can be carried out according to the pressure rise method.
Special versions::High pressure, low pressure
Typical applications::Water or oil leak testing
Operating mode (standard)::Single-channel operation
Operating mode (option)::synchronous two-channel operation
Signal Processing::24-bit A / D converter, real-time
Valve types::sliding valve or poppet valve
Flow control::Microcontroller 16-bit μC / 40 MHz
Pressure sensors 1::Differential pressure sensor
Pressure sensors 2::gauge pressure sensor
Measurement ranges::± 500 Pa / ± 5,000 Pa
Result units 1::Pa, hPa, PSI, Pa/s, hPa/s, mbar • l/s
Result units 2::ml/min, ml/h, l/min, l/h
Result units 3::mmHg, mmWs, Torr
Pressure ranges 1::-1 bar, +/- 5 mbar, +/- 50 mbar
Pressure ranges 2::0,2 bar, 1 bar, 6 bar, 10 bar, 20 bar
Pressure ranges 3::30 bar, 400 bar (high pressure)
Standard test modes::pressure decay (drop)
Test modes Optional 1::sealed component, dynamic pressure
Test modes Optional 2::Pressure steps, pressure rise
Test Options 1::Prefilling, program series